Review Article - (2020) Volume 6, Issue 1
1Department of Cardiology, University Hospital Zurich, Switzerland
2Heart Center Lucerne, Luzerner Kantonsspital, Switzerland
3CardioVascular Center Zurich, Hirslanden Klinik Im Park, Zurich, Switzerland
Received Date: December 02, 2019; Accepted Date: February 03, 2020; Published Date: February 10, 2020
Citation: Keller LS, Toggweiler S, Sutsch C, Obeid S, Tanner FC, et al. (2020) Outcomes after Implantation of the ACURATE Neo and Portico Transcatheter Heart Valves for the Treatment of Severe Aortic Stenosis. Interv Cardiol J Vol.6 No.1:88. doi: 10.36648/2471-8157.6.1.88
Copyright: © 2020 Keller LS, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Introduction: Among others the ACURATE neo and the Portico self-expanding transcatheter heart valves are widely used to treat severe aortic stenosis. So far, a direct comparison of the hemodynamics of these two valves is lacking. We want to fill this gap by a retrospective analysis of hemodynamic performance and the occurrence of new conduction disturbances.
Methods and results: Prospectively collected data at the University Hospital Zurich and the Heart Center Lucerne between December 2012 and April 2018 were analyzed. A total of 318 consecutive patients undergoing implantation of an ACURATE neo or a Portico valve formed the study population. The ACURATE neo was implanted in 144 patients (44% male) and the Portico in 174 patients (47% male). Patients receiving the ACURATE neo were older (82 ± 6 vs. 80 ± 7, p=0.03), had a higher LVEF (58 ± 12% vs. 54 ± 14%, p= 0.01) and a higher mean transvalvular pressure gradient at baseline (49 ± 17 vs. 41 ± 17 mmHg, p<0.001). There was no difference in annular size between the two groups (a perimeter of 75.3 ± 8.6 vs. 75.4 ± 5.2 mm, p=0.94).
Incidence of > mild paravalvular leak was low in both groups (3.4% in Portico vs. 5.6 % in ACURATE neo, p=0.42) at 30 days. The mean transvalvular pressure gradient after implantation of ACURATE neo was comparable to Portico (7 ± 4 mmHg vs. 8 ± 4 mmHg, p=0.05). New pacemaker insertion was significantly less frequent in the ACURATE neo group (2.5% vs. 10.9%, p=0.01).
Conclusion: Hemodynamic outcomes between the intra-annular Portico and the supra-annular ACURATE neo valve were similar with low transvalvular pressure gradients observed after implantation of both valves. Pacemaker rates after ACURATE neo implantation was lower
Transcatheter aortic valve implantation (TAVI); Self-expandable valve; ACURATE neo-Portico
Transcatheter aortic valve implantation (TAVI) has been established as the treatment of choice in a broad patient population with aortic stenosis ranging from low-risk to inoperable patients [1,2]. Currently, the main limitations of TAVI are paravalvular leaks (PVL) and the need for new pacemaker implantation [3,4]. Furthermore, long-term outcome data (beyond 10 years) are currently missing.
The ACURATE neo (Boston Scientific, Marlborough, MA, USA) is a second-generation self-expanding, non-repositionable, supra-annular, porcine pericardial leaflet valve system. The Portico (Abbott, Abbott Park, IL, USA) is a second-generation self-expanding, repositionable, intra-annular, bovine pericardial leaflet valve system.
Data directly comparing the two valves are currently missing. Given the two different concepts (intra- vs. supra-annular valve fixation; repositionable vs. non-repositionable) such a comparison is of interest, as the design of the valve potentially impacts hemodynamics and conduction disturbances.
The aim of the present study was therefore to compare the hemodynamic performance and the incidence of permanent pacemaker implantation of the ACURATE neo and the Portico valve.
Study design and population
Consecutive patients with severe aortic stenosis, defined as an aortic valve area of <1.0 cm2 or a mean transvalvular pressure gradient >40 mmHg or maximal velocity of >4.0 m/s, treated at the University Hospital Zurich and Heart Center Lucerne between December 2012 and April 2018, using either a Portico or ACURATE neo valve prosthesis, were included in the present analysis. Patients undergoing TAVI with other valve types than Portico and ACURATE neo were not eligible for inclusion.
Patients gave written informed consent to the procedure and to data collection, approved by the local ethics committees (Swiss TAVI Registry (PB_2016-00394)).
Procedural characteristics
Pre-procedural work-up included a transthoracic echocardiogram and a gated-CT scan. Transthoracic echocardiography studies were performed by experienced certified personnel. Studies were analysed according to the American Society of Echocardiography (ASE) and the European Association of Echocardiography (EAE) recommendations and performed using commercially available ultrasound systems (Philips iE33 or Epic, Philips Healthcare, Andover, MA, USA; GE Vivid 7 or E9 or E95, GE Healthcare, Milwaukee, WI, USA). Basic echocardiographic parameters were retrieved from baseline echocardiographic reports. LV ejection fraction (LVEF) was measured by Simpson’s biplane method.
The vast majority of transcatheter aortic valve implantation procedures were performed via the transfemoral approach (95.4% in Portico and 95.1% in ACURATE neo) in local anaesthesia and under fluoroscopic guidance only. Pre-dilatation of the native aortic valve was done in the majority of cases (86.2% in Portico and 95.1% in ACURATE neo). At 30 days a transthoracic echocardiogram and a clinical follow-up were performed.
Statistical analysis
Continuous variables are expressed as mean ± SD or medians with interquartile ranges (IQR), and were compared using 1-way ANOVA, Student’s t-test, Kruskal-Wallis or Mann-Whitney tests, as appropriate. Categorical data are presented as frequency (percentages) and were compared using the Fisher exact or the chi-square test. All statistical analyses were performed with SPSS 22 and p-values < 0.05 were considered as statistically significant.
Study endpoints
At 30 days, mean transvalvular pressure gradient, severity of paravalvular leaks and incidence of permanent pacemaker implantation were recorded. Further endpoints analyzed at baseline and at 30 days included functional capacity, as defined by New York Heart Association classification, and changes in left ventricular ejection fraction.
A total of 318 patients (45% male, age 81 ± 7) were included. The Portico was used in 174 patients (47% male, age 80 ± 7) and the ACURATE neo in 144 patients (44% male, age 82 ± 6). The ACURATE neo population had a higher STS PROM (4.6 ± 4.2 vs. 4.3 ± 2.7) than the Portico group.
Baseline clinical and echocardiographic data are listed in (Tables 1 and 2). Outcome data after transcatheter aortic valve implantation with the Portico and ACURATE neo valve systems are listed in (Tables 3 and 4).
Characteristics | Portico (N=174) | ACURATE neo (N=144) | p-value |
---|---|---|---|
Age - years | 80.3 ±7.2 | 82.0 ± 6.2 | 0.028 |
Male sex (%) | 82/174 (47.1) | 63/144 (43.8) | 0.573 |
STS Risk Score | 4.3 ± 2.7 | 4.6 ± 4.2 | 0.435 |
Previous MI (%) | 15/174 (8.6) | 13/144 (9.0) | 1 |
PTCA/Stenting (%) | 43/174 (24.7) | 20/144 (13.9) | 0.017 |
CABG (%) | 12/174 (6.9) | 6/144 (4.2) | 0.338 |
Atrial fibrillation (%) | 56/174 (32.2) | 277138 (19.6) | 0.014 |
NYHA ³II (%) | 135/174 (77.6) | 136/142 (95.8) | <0.001 |
Cardiovascular risk factors (%) | |||
Diabetes mellitus | 41/174 (23.6) | 29/144 (20.1) | 0.499 |
Hypertension | 126/174 (72.4) | 118/144 (81.9) | 0.047 |
Table 1: Baseline characteristics. Patients receiving an ACURATE neo valve were older and in worse functional class, while patients receiving the Portico valve were more often in atrial fibrillation and previous stenting/PCTA were more frequent.
Characteristics | Portico (N =174) | ACURATE neo (N =144) | p-value |
---|---|---|---|
LVEF - % | 54.0 ± 14.1 | 58.3 ± 11.8 | 0.004 |
Mean transvalvular pressure gradient - mmHg | 40.6 ± 16.5 | 48.7 ± 17.3 | <0.001 |
Perimeter Aortic Annulus - mm | 75.3 ± 8.6 | 75.4 ± 5.2 | 0.941 |
Table 2: Echocardiographic and CT data at baseline. Left ventricular ejection fraction and mean transvalvular pressure gradient were lower in the Portico group.
Characteristics | Portico (N =174) | ACURATE neo (N =144) | p-value |
---|---|---|---|
Mortality (%) | 8/174 (4.6) | 5/144 (3.5%) | 0.778 |
Myocardial Infarction (%) | 1/174 (0.6) | 0/144 (0) | 1 |
Major/Life-threatening Bleeding (%) | 15/174 (8.6) | 6/144 (4.2) | 0.12 |
Second valve Implantation | 3/174 | 1/143 | 0.63 |
NYHA ≥ II (%) | 84/155 (54.2) | 48/133 (36.1) | 0.003 |
New PM Implantation (%) | 19/155 (10.9) | 3/117 (2.5) | 0.006 |
Table 3: Clinical outcomes at 30 days. New pacemaker implantation was more frequent in the Portico group and the functional outcome was worse. However there were no significant differences in terms of mortality, myocardial infarction, major bleeding and need of a second valve between the two groups
Characteristics | Portico (N=174) | ACURATE neo (N =144) | p-value |
---|---|---|---|
Paravalvular Leakage>mild (%) | 6/174 (3.4) | 8/144 (5.6) | 0.417 |
Mean transvalvular pressure gradient - mmHg | 7.7 ± 3.5 | 7.1 ± 3.9 | 0.05 |
DLVEF (Baseline to 30d) - % (total no.) | +2.4 (172) | +2.3 (136) | 0.955 |
Table 4: Hemodynamic Outcomes at 30 days. Paravalvular leakage, mean transvalvular pressure gradient and ΔLVEF were similar in patients receiving the ACURATE neo and the Portico valve.
Hemodynamic outcomes
Mean trans-valvular pressure gradient decreased from 40.6 ± 16.5 to 7.7 ± 3.5 mmHg in the Portico and from 48.7 ± 17.3 to 7.1 ± 3.9 mmHg in the ACURATE neo group (p < 0.001 for both valves compared to baseline). A para-valvular leak >mild was rare in both groups (3.4% in Portico vs. 5.6% in ACURATE neo, p=0.42) (Figure 1).
Clinical outcomes
Procedural outcomes were comparable between the two groups. Mortality at 30 days in the Portico and the ACURATE neo group was 4.6% vs. 3.5% (p=0.78) and major bleeding complications (major or life-threatening bleedings) occurred in 8.6% vs. 4.2% (p=0.12). Myocardial infarction was reported in only one patient (0.6% vs. 0% (p=1.0)). Implantation of a second valve was required in 4 patients (3 in the Portico group and 1 in the ACURATE neo group, p=0.63). The incidence of permanent pacemaker implantation was significantly higher in the Portico group compared to the ACURATE neo group (10.9% vs. 2.5%, p=0.01). A logistic regression model corrected for age, history of hypertension, atrial fibrillation and previous PTCA/Stenting showed that the Portico valve was associated with a five-fold increased risk of new permanent pacemaker implantation within 30 days (OR 5.184 (1.465-18.344); p=0.01).
At 30 days functional recovery was worse in the Portico group (NYHA ≥ II 54% vs. 36%, p=0.01). At baseline functional class NYHA ≥ II was more frequent in the ACURATE neo group, reflecting a better recovery of the ACURATE neo group.
The main finding of this retrospective comparison between the two valve systems ACURATE neo and Portico was a significantly lower pacemaker rate after implantation of the ACURATE neo. There were no differences in hemodynamic outcome in terms of paravalvular leakages and transvalvular pressure gradients at 30 days follow-up. New permanent pacemaker rates after TAVI using earlier-generation valve systems ranged between 5% and 12% for balloon-expandable and between 24% and 33% for self-expanding devices [5]. With more precise implantation techniques resulting in prostheses being implanted more aortic, and thanks to improvements in prosthesis design, pacemaker rates have come down to 6.5% for balloon-expandable valves [2]. Pacemaker rates for self-expanding valves on the other hand, tend to remain higher [6, 7].
We found in our cohorts low rates of new pacemaker implantation – both for the Portico and the ACURATE neo valve. A new permanent pacemaker was required in 10.9% of patients receiving the intra-annular valve system Portico, whereas a new pacemaker rate of 18.7% at 30 days was reported in the PORTICO-I trial [6]. An even lower pacemaker rate was found for the cohort receiving the ACURATE neo valve (2.5%), achieving results as with balloon-expandable valves. This is lower than reported rates for other supra-annular valve systems- e.g. the Evolute valve (13.3%) [8]. Also the rates are lower than reported in other studies using the ACURATE neo valve (8.2% and 10.2%) [9]. The reasons for this are unknown and are speculative - however, routine cessation of negative dromotrope medication as well as aiming at an aortic implantation may explain the findings.
The current Portico valve does not feature a sealing cuff. We learned from our own unpublished data, that aggressive oversizing when using the Portico valve resulted in very low paravalvular leak rates. This has become routine practice at our institution. The downside of this practice may be a slightly higher pacemaker rate and may explain the differences observed between the ACURATE neo and the Portico valve in our cohort [10].
Our data do not support the suggested lower transvalvular pressure gradients when using supra-annular as compared to intra-annular prostheses. While both designs may have their advantages, vascular access, the presence of coronary artery disease as well as center experience may play an important role on valve choice. Furthermore, with expanding indications for TAVI and use of TAVI in younger patients, the suitability for valvein- valve procedures should also be considered.
This study is a retrospective analysis, generated from real-world observational data. The local Heart Team decision-making process cannot be captured and may be a source of bias. This study focuses on short-term outcomes (30 days observational period) consequently conclusions on long-term hemodynamic differences between the two valve systems cannot be made.
Dr Nietlispach Consultant and Proctor for Abbott Vascular and Edwards Lifesciences;
Dr Toggweiler Consultant and proctor for Boston Scientific, NVT GmbH and Abbott Vascular, has received institutional research grants from Boston Scientific and Fumedica AG and holds equity in Hi-D Imaging AG